Your enquiry has been submitted
2X Urea Denaturing Gel Loading Buffer
Technical Specification
| Product Name | 2X Urea Denaturing Gel Loading Buffer |
|---|---|
| Product Code | ML176-1ML |
| Kit Packing | 1ml |
| Composition | 2X Urea Denaturing Gel Loading Buffer consists of 90% Formamide, EDTA, Bromophenol Blue and Xylene Cyanol as tracking dyes. |
| Appearance | Blue color solution |
| Clarity | Clear and free of particles |
| Suitability Test | This solution has been tested and is suitable for use with polyacrylamide urea gel (denaturing) and non-denaturing agarose gel electrophoresis of nucleic acids |
| Storage conditions | 2-8°C. Under recommended condition, the reagent is stable for 6 months. |
| Expiry Date | 02/2025 |
Description of the product
2X Urea Denaturing Gel Loading Buffer is recommended for use with polyacrylamide urea gel (denaturing) agarose gel electrophoresis of nucleic acids. 2X Urea Denaturing Gel Loading Buffer is suitable for use in loading nucleic acid samples in polyacrylamide urea gel (denaturing) and non-denaturing agarose gel electrophoresis of nucleic acids.
Procedure
Sample preparation and Loading on Agarose Gels:
- Add 25µl of ML176 to 25µl of sample and Mix properly.
- Load 10µl of prepared sample in 1% AGE gel
- Run the gel in 1X TAE/ TBE Buffer till the dye front reaches three-fourth of the gel.
NOTE: Prestaining of the gel with Ethidium bromide is recommended
Sample Preparation and Loading on Urea PAGE:
- Add 25µl of ML176 to 25µl of sample and Mix properly.
- Loading can be done with or without heating.
- Heating can be done at 95°C for 5 minutes.
- Load 10µl of prepared sample in Urea-PAGE gel
- Run the gel in 1X TBE Buffer at 130V for 55mins.
NOTE: For Urea denaturing PAGE, post staining of the gel with Ethidium bromide is recommended.
Warning and Precautions
Not for Medicinal Use. Read the SDS carefully before beginning the protocol. Wear protective gloves/protective clothing/eye protection/face protection. Follow good clinical laboratory practices while handling clinical samples. Standard precautions should be followed as per established guidelines. Safety guidelines may be referred in safety data sheets of the product.
Performance and Evaluation
Performance of the solution is expected when the solution is stored at recommended temperature and within the expiry period.
Safety Information
The 2X Urea Denaturing Gel Loading Buffer is for laboratory use only, not for drug, household or other uses. Take appropriate laboratory safety measures and wear gloves and safety goggles when handling. Not compatible with disinfecting agents containing bleach. Please refer the Safety Data Sheet (SDS) for information regarding hazards and safe handling practices.
Technical Assistance
At HiMedia we pride ourselves on the quality and availability of our technical support. For any kind of technical assistance, mail at mb@himedialabs.com.
| Product Name | 2X Urea Denaturing Gel Loading Buffer |
|---|---|
| SKU | ML176 |